Leica Biosystems · Model RM2245
High precision specimen feeding: High precision specimen feeding via stepper motor provides consistent section quality for sensitive staining methods. Large 70mm vertical stroke: The large 70mm vertical stroke allows sectioning of super mega cassettes and also supports safer specimen changing due to the greater distance to the knife edge. Adjustable handwheel balance system: Patented adjustable handwheel balance system protects the specimen because the object head does not drop into the knife when sectioning is stopped. Precision orientation system: The optional 8° X/Y precision orientation system with calibrated controls helps to quickly orient already cut specimens (re-cuts). The system then rapidly returns the exact zero position, which is marked by two red indicators.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Thick-thin sectioning, chatter, ribbons that will not form, scoring lines.
Root cause: Dull or nicked blade, incorrect clearance angle, a loose specimen / blade clamp, or worn feed mechanism.
Symptoms: Section thickness inconsistent or drifting, coarse feed will not advance / retract.
Root cause: Worn micro-feed mechanism, a stepper / clutch fault (motorised), or backlash.
Symptoms: Handwheel binds, the brake will not hold, or motorised sectioning stalls.
Root cause: Worn bearings, a brake fault, or the drive motor / belt (motorised).
Symptoms: Chamber or specimen head will not reach temperature, heavy frost.
Root cause: Refrigeration fault, a temperature-sensor fault, or a failed defrost cycle.
Symptoms: Block moves during sectioning, orientation will not lock.
Root cause: Worn specimen clamp or orientation mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Blade holder / clamp assembly | Major service part | $$ | Moderate |
| Micro-feed mechanism | Major service part | $$$ | Difficult |
| Handwheel bearings / brake | Major service part | $$ | Difficult |
| Cryostat refrigeration / temp sensor (cryostat) | Major service part | $$$ | Difficult |
| Specimen clamp / orientation head | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.