Leica Biosystems · Model ASP6025
Simple, easy-to-use LCD touch screen with an intuitive display and easy to follow icons helps reduce operator error Accurate diagnosis is facilitated when samples are optimally processed. A built in densitometer automatically monitors ethanol concentration ensuring that specimens are processed in the proper dilution. Advanced Specimen Safety - built-in battery backup system protects tissue samples in the event of a power outage Quickly and easily view specimens through the retort viewing window to see that your specimens are fully covered during processing Leica RemoteCare Service offers remote monitoring to help identify and eliminate problems before tissue damage occurs Unique pre-melt paraffin station, the system helps avoid delays and ensures that paraffin is always available and melted prior to exchanging used paraffin Eliminate the need to mix and prepare dilutions and spend less time maintaining the processor all while ensuring optimal specimen processing Ready to Use (RTU) bottles reside in an easy to access drawer ergonomically designed to minimize spills and injuries The remote drain system easily and safely drains hot used paraffin and lowers the risk of a paraffin spill.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Stations not filling or draining, retort not filling (processor), transport arm mis-places baskets, error alarm halts the run.
Root cause: Worn pump / valve, a blocked line, a level-sensor fault, or a transport-drive / gripper fault.
Symptoms: Pressure / vacuum not achieved, poor infiltration, seal alarm.
Root cause: Worn retort door seal, a vacuum / pressure-pump fault, or a valve.
Symptoms: Slides over- / under-stained, uneven, station-time drift.
Root cause: Exhausted reagents, wrong programme, incorrect agitation, or a timing / arm fault.
Symptoms: Paraffin not molten or too hot, oven not at temperature, bath alarm.
Root cause: Heater or temperature-sensor fault.
Symptoms: Reagent fumes in the room, filter-change overdue.
Root cause: Saturated carbon filter or a fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reagent / paraffin pump & valves | Periodic replacement | $$$ | Difficult |
| Retort / reagent-vessel seals | Periodic replacement | $$ | Fairly easy |
| Transport-arm drive / gripper (stainer) | Periodic replacement | $$$ | Difficult |
| Bath / oven heater & temperature sensor | Major service part | $$ | Moderate |
| Activated-carbon fume filter | Frequent wear item | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.