IBIon Beam Applications SA · Model adaPTinsight
Image Guided Therapy Console.
Intended use: An automated device that is typically mounted on the gantry of a linear accelerator and intended to produce digital images of x-rayed anatomical landmarks to guide radiation treatment (e.g., tracking/targeting tumours). The device may be a digital imaging panel (e.g., of silicon) on robotic arms; it may also have an x-ray source to generate higher quality images.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Temporary deactivation of collision detection for Forte Robotic Patient Positioning System occasionally experiences issues
Safety Parameters Verification Mechanisms can be deactivated in clinical runtime
Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances
There is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.
The patient positioning system can perform motions in unexpected directions, creating the potential for crushing of the patient between the patient positioning devices.
IBA is initiating this recall to address an issue identified with AdaPTinsight software and to reduce risk related to this problem.
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du