Symptoms: Breathing-circuit / absorber leak test fails, machine not ready for use.
Root cause: Breathing-circuit / absorber leak, a worn APL valve or bag arm, or a CO2-absorber canister seal.
Symptoms: O2 flow not matching the setting, flowmeter float sticking, no O2 flush.
Root cause: Worn flowmeter (rotameter) or a flush-valve fault.
Symptoms: Agent not vaporising correctly, output does not match the dial, back-bar leak.
Root cause: Vaporiser seating / O-ring, or an internal vaporiser fault.
Symptoms: Set tidal volume not delivered, bellows will not fill, PEEP not held.
Root cause: Bellows leak, drive-gas valve, or a flow-sensor fault.
Symptoms: Waste-gas scavenging alarm or not evacuating.
Root cause: Scavenging flow / connection or a receiver-valve fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| APL valve / breathing-circuit gaskets | Periodic replacement | $$ | Fairly easy |
| CO2 absorbent / canister seals | Consumable | $ | Easy |
| O2 flowmeter (rotameter) | Major service part | $$ | Moderate |
| Vaporiser O-ring / back-bar seal | Periodic replacement | $ | Fairly easy |
| Ventilator bellows (Versa) | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio