Huntleigh · Model Sonicaid FM800
Antepartum CTG analysis- Provides and objective, numerical and consistent assessment of the CTG. Clinicians can rely on the worlds largest database of collective experience to support their own visual assessment. Intrapartum trend- Measures changes over time in the FHR. For use during the 1st stage of labor, clinicians are quickly alerted to changes in short term variability, baseline rate and deceleration size.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No fetal heart rate displayed, frequent dropouts, poor trace.
Root cause: Ultrasound transducer crystal degraded or cracked, insufficient gel / poor placement, or a damaged transducer cable.
Symptoms: Uterine activity flat or noisy, will not zero.
Root cause: Toco transducer strain gauge, belt tension, or zero-reference fault.
Symptoms: Monitor runs but no paper output.
Root cause: Out of paper, paper tray not fully pushed in, or a thermal print-head fault.
Symptoms: Error message, monitoring stops.
Root cause: Internal fault - the manual directs the user to stop, note the message and return the unit to Huntleigh.
Symptoms: MSpO2 shows no value when the sensor is connected.
Root cause: Oximetry module not connected to the FM800, or a sensor fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
NBP module calibration every 12 months.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Ultrasound (FHR) transducer | Periodic replacement | $$ | Easy |
| Toco (contraction) transducer | Periodic replacement | $$ | Easy |
| Thermal print head and chart paper | Periodic replacement | $ | Fairly easy |
| Fetal event marker | Periodic replacement | $ | Easy |
| NBP cuff and hose NBP module calibration every 12 months. | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.