Hospira · Model LifeCare PCA 2
Abbott Lifecare PCA 2 4100 Infusion Pump IV
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Screw rotation error occurring when the pump detects a mismatch between the number of steps given to the motor driving delivery and the number of steps the motor actually took. When this malfunction occurs, the pump will alarm and infusion stops.
The infusion pump has a clear plastic door that, due to its design, allows for excessive flexing. This in turn permits unauthorized access to the vial of medication.
Reports of leaking during filling and administration.
Fire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.
Incorrect dosage labels created: Customers have experienced independent changes in drug concentration (mg/mL to mcg/mL and vice versa) and expiration dates, and the addition of trailing zeros to the barcodes.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.