Hettich · Model MIKRO 22 / 22 R
Three complete running programs can be stored and recalled. Automatic lid locking, lid with see-through glass for speed control during centrifugation. Clearly arranged, water-protected foil keypad for entering the parameters. Stand-by mode (200R) maintains temperature / auto refrigeration shut-off after set time or when lid open. Brushless drive guarantees exceptionally smooth running and speed stability. Clearly arranged display: During centrifugation the actual values of the parameters are digitally displayed. Automatic rotor recognition. Speed (RPM): Input in increments of 10. Relative Centrifugal Force (RCF): Input in increments of 10. Temperature control in refrigerated models within a range of -20°C to +40C maintenance free frequency induction drive. Imbalance switch-off. Running time (t): Input in minutes and seconds. 2 year warranty.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Error code, run will not begin, or aborts shortly after start.
Root cause: Set speed exceeds the rotor maximum, an imbalanced load, the lid / door not locked, an unapproved rotor, or a drive / control fault.
Symptoms: Slow spin-up, rumble, or vibration during a run.
Root cause: Load imbalance, worn drive / motor bearings, a worn drive coupling, or (older units) worn motor brushes.
Symptoms: Temperature high during a run, or slow to pre-cool.
Root cause: Compressor, condenser fan, low refrigerant, or a temperature-sensor fault.
Symptoms: Lid-lock error, will not open after a run, or will not lock to start.
Root cause: Lid-lock solenoid / switch fault, or a run still winding down.
Symptoms: Visible corrosion, cracks, or a bucket that will not swing; imbalance trips.
Root cause: Rotor / bucket ageing - a safety-critical wear item with a use-life limit.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Log rotor runs; retire per the manufacturer's rotor-life programme.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive / motor bearings & coupling | Major service part | $$$ | Difficult |
| Lid-lock solenoid / switch | Major service part | $$ | Moderate |
| Speed sensor / control board | Major service part | $$ | Moderate |
| Condenser fan / refrigeration components | Major service part | $$$ | Difficult |
| Rotor / bucket set & O-rings Log rotor runs; retire per the manufacturer's rotor-life programme. | Major service part | $$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.