Haemonetics · Model OrthoPAT
Dynamic Disk technology that changes size according to blood volume loss and provides a consistent hematocrit in the reinfusable blood product. Intraoperative and postoperative salvage in a single disposable set. Easy loading disposable set. Full automation that requires minimal operator interaction. Continuous display of the volume of collected fluid. Small and highly portable device that mounts to an IV pole and is easy to transport, clean, and store.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "Centrifuge speed error", "Check disk/retaining ring placement".
Root cause: Disk not loaded, retainer ring missing, or the header arm not properly engaged with the disk header.
Symptoms: System will not build or hold pressure, pressure alarms.
Root cause: Damaged or missing centrifuge D-ring, or debris in the D-ring groove.
Symptoms: "Cannot process on battery" when switched to battery power mid-process.
Root cause: Battery power does not support processing by design; or an aged battery that will not recharge in 4 h.
Symptoms: Fluid in the centrifuge well, contamination alarm.
Root cause: Overflow / capacity exceeded, or a disposable-set breach.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-procedure consumable; confirm recall status.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable Dynamic Disk set Single-procedure consumable; confirm recall status. | Consumable | $ | Easy |
| Centrifuge D-ring | Periodic replacement | $ | Fairly easy |
| Retainer ring | Periodic replacement | $ | Easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Header arm assembly | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
2 Models Model 1: OrthoPAT devices with no centrifuge drain spill containment system may have potential for fluid contamination leading to potential spark or smoke may occur are recalled Model 2: OrthoPat equipped with a spill collection drainage system updated with revised Instructions for Use and Quick reference Guide Addendum