GenesisBPS · Model Blood Collection Monitor
Electronically monitors and measures blood flow and volume. Automatically clamps tubing when target volume is achieved. Rechargeable battery provides up to 50 hours of service. Automatic weight calibration. Automatically tares collection bag. Audible & visual alarms. Mix reminder alarm. Slow bleed alarm. Conserves battery with automatic shut off after 30 minutes of idle time.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Displayed volume does not match the actual collected weight, will not tare / zero.
Root cause: Load-cell drift or damage, a binding cradle, or a calibration fault.
Symptoms: The rocker does not agitate the bag, uneven or no motion, motor noise.
Root cause: Worn rocker drive / motor or linkage.
Symptoms: Auto clamp does not close at target volume or on slow flow, flow-rate alarm.
Root cause: Tubing-clamp solenoid / mechanism, or a flow-detection fault.
Symptoms: No or quiet end-of-collection or slow-flow alarm.
Root cause: Speaker or audio-board fault.
Symptoms: Short battery life, will not power on.
Root cause: Aged battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate with certified weights after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Load cell Recalibrate with certified weights after replacement. | Major service part | $$ | Difficult |
| Rocker drive / motor | Major service part | $$ | Moderate |
| Tubing-clamp solenoid / mechanism | Periodic replacement | $$ | Moderate |
| Battery | Periodic replacement | $$ | Fairly easy |
| Alarm speaker | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.