GE HealthCare · Model XR 6000 - Silhouette
Exceptional images. A Dual Focal Spot tube (0.6/1.2 mm) allows clear visualization of large and small structures. Efficient workflow. An operator-friendly and intuitive console minimizes staff training. An anatomical programmer supports high image quality at low dose and virtually eliminates retakes Ergonomics and patient comfort. The column automatically tracks with the detector, allowing fast exam setup. A low table height (69.5 cm) makes patient positioning easy. Ease of installation. A free-standing floor-mount design minimizes room preparation and allows installation in as little as one day. An integrated under-table generator creates a compact footprint and allows flexibility to fit rooms as small as 5.5 m by 4 m. Reliability. High-strength structural components ensure long life and reliable performance. The design allows easy service access to the generator and power distribution unit. Systems come backed by a network of more than 750 Field Service Engineers throughout Europe, Middle East and Africa, plus world-class parts delivery.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, generator error / fault code, images consistently under- or over-exposed.
Root cause: HV generator / inverter fault, an exposure-switch / contactor fault, an HT-cable fault, or a tube-load / thermal interlock.
Symptoms: Output does not match the console, techniques drift, generator calibration fails.
Root cause: Inverter / HV-tank fault, a filament-supply fault, or a calibration issue.
Symptoms: Auto-exposure images consistently over- or under-exposed, chambers not tracking.
Root cause: AEC ion chamber(s) or the AEC calibration.
Symptoms: Column, wall stand, table or bucky will not track / detent; collimator light out; grid errors.
Root cause: Drive / brake / detent hardware, the collimator lamp / drive, or a bucky-motor fault.
Symptoms: Exposures inhibited after a series, tube-heat alarm.
Root cause: Tube housing heat limit reached or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| HV generator / inverter components | Major service part | $$$$ | Expert-level |
| HT cables | Major service part | $$$ | Difficult |
| X-ray tube assembly | Major service part | $$$$ | Expert-level |
| AEC ion chamber | Major service part | $$$ | Difficult |
| Collimator lamp / drive assembly | Periodic replacement | $$ | Moderate |
| Bucky / grid motor | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio