GE HealthCare · Model Proteus XR/i
Flexible and User-Friendly • The broad clinical flexibility of Proteus XR/i system enables you to undertake a wide variety of radiographic studies, including tomography as an option. • The floor-mounted tube stand, the floating Bucky table and the electromagnetic brakes make for quick and easy patient positioning. • The elevating table is equipped with central foot switches. The wall stand with oscillating Bucky allows fast and accurate positioning. Image Quality • The Proteus XR/i benefits from GE Healthcares uncompromising quality standards that ensure very high uptime ratings. • The high-power, high-frequency generator with its very short minimum exposure time provides excellent reproducibility that results in consistent image quality. • Automatic Exposure Control (AEC) reduces the number of retakes, and the Anatomical Programmer (APR) accommodates a full range of studies. • The system can be equipped with the Dose Area Product (DAP) measurement device with printer as an option. Reliability and Ease of Use • To enhance your productivity and ensure maximum uptime, the system is equipped with a single-phase or three-phase high-frequency generator (50/64 kW) that is suitable for almost all applications. • Ergonomic and easy-to-use Touch screen console • A very wide range of accessories such as lateral cassette holder, compression band, hand grips, etc
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Tube won't fire, heat-capacity alarm.
Root cause: Filament wear, cooling failure.
Symptoms: Image noise, artifacts, low SNR.
Root cause: Flat-panel / sensor aging.
Symptoms: Tube temp alarm, duty-cycle limited.
Root cause: Pump/radiator or coolant loss.
Symptoms: Exposure fails, kV/mA out of range error.
Root cause: HV tank or generator board fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Major R&R; track mAs history.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| High-voltage generator | Major service part | $$$$ | Expert-level |
| Workstation RAID / GPU | Periodic replacement | $$$ | Difficult |
| X-ray tube insert Major R&R; track mAs history. | Major service part | $$$$ | Expert-level |
| Flat-panel detector | Major service part | $$$$ | Expert-level |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Potential breakage of the steel cable which supports the bucky device in certain wall stands of Proteus XR/a X-ray imaging systems. A fall of a bucky while the system is in use could result in an injury to a patient or operator.
The warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
The actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.