GE HealthCare · Model Lunar Prodigy Primo
Proven bone densitometer - The Lunar Prodigy® bone densitometer features direct-digital technology that eliminates the magnification effects of traditional wide-angle fan beam densitometers. With more than 6,000 systems installed worldwide, Prodigy is one of the most trusted fan beam densitometers available. OneVision - With OneVision you and your referring physicians receive a single, consolidated report that combines the complete risk assessment analysis rather than receiving multiple assessment reports. Dual-energy Vertebral Assessment-DVA - DVA expands the clinical applications available for the Lunar Prodigy Primo bone densitometer. DVA provides rapid, dual-energy images of the AP and lateral spine, allowing clinicians to visually assess vertebral deformations. DualFemur - The DualFemur feature automatically measures both left and right femurs in one fast scan, and improves accuracy by identifying the femur with the lowest density. The 30? improvement in precision enhances the ability to monitor response to therapy. Patient Reports - The Lunar Prodigy Primo patient report displays T-scores and Z-scores along with the fracture risk assessment graph, based on the World Health Organization (WHO) criteria to define osteoporosis, and key diagnostic information. This makes interpretation of results and fracture risk assessment seamless. Total-body-body composition - Body composition scans with DXA provide precise and accurate data on bone and tissue composition, including bone mineral density (BMD), lean and fat tissue mass, and %fat. They provide both total body data and regional results. Benefits - Effective use of serial DXA measurements for monitoring changes in BMD requires the minimization of precision error. The Lunar Prodigy® Primo has been demonstrated to have a precision error that easily meets the ISCD standards for all regions. Technology - The Lunar Prodigy Primo utilizes a direct-digital array detector and narrow-angle fan-beam technology to enhance dose efficiency and achieve excellent precision and patient throughput in spine, femur and total body measurements.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Daily spine / block phantom BMD trends out of control limits, or calibration fails.
Root cause: X-ray source output drift, detector degradation, or a calibration-drum / reference fault.
Symptoms: The scanning arm or table will not move smoothly, position error, scan aborts, image geometry wrong.
Root cause: Drive motor, belt, or an encoder / limit-sensor fault.
Symptoms: Exposure inhibited, kV error, no counts / very noisy scan.
Root cause: X-ray tube / HV supply, or the filter-wheel / calibration-drum.
Symptoms: Missing scan lines, non-uniform image, detector-error message.
Root cause: Detector element / electronics degradation.
Symptoms: Analysis software hangs, scans will not archive or transfer.
Root cause: Workstation PC or software fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| X-ray tube / HV supply | Major service part | $$$$ | Expert-level |
| Detector assembly | Major service part | $$$$ | Difficult |
| Scanning-drive motor / belt / encoder | Major service part | $$$ | Difficult |
| Calibration drum / filter wheel | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.