GE HealthCare · Model Lunar Achilles InSight
Automated and self-contained - The standard system is an automated and self-contained unit with an internal printer. The on-board memory stores up to 300 patient results, which you can review or print at your convenience. Printouts are easily attached to permanent patient records. DICOM - DICOM (Digital Imaging and Communications in Medicine) is a standard for transferring digital media images and associated information between devices. The DICOM standard specifies information that vendors must supply about their product, when the product claims to be DICOM standard compliant. Lunar’s DICOM Conformance Statement defines what DICOM protocols our product includes and the details it uses in communication with an existing Picture Archiving Communications System (PACS) or Radiology Information Server (RIS).
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Daily spine / block phantom BMD trends out of control limits, or calibration fails.
Root cause: X-ray source output drift, detector degradation, or a calibration-drum / reference fault.
Symptoms: The scanning arm or table will not move smoothly, position error, scan aborts, image geometry wrong.
Root cause: Drive motor, belt, or an encoder / limit-sensor fault.
Symptoms: Exposure inhibited, kV error, no counts / very noisy scan.
Root cause: X-ray tube / HV supply, or the filter-wheel / calibration-drum.
Symptoms: Missing scan lines, non-uniform image, detector-error message.
Root cause: Detector element / electronics degradation.
Symptoms: Analysis software hangs, scans will not archive or transfer.
Root cause: Workstation PC or software fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| X-ray tube / HV supply | Major service part | $$$$ | Expert-level |
| Detector assembly | Major service part | $$$$ | Difficult |
| Scanning-drive motor / belt / encoder | Major service part | $$$ | Difficult |
| Calibration drum / filter wheel | Major service part | $$$ | Difficult |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio