GE HealthCare · Model 3SC-RS
Ultrasound Transducer GE's 3SC-RS sector array phased array ultrasound probe transducer. A cardiac probe, with a frequency range of 1 - 4 MHz., this versatile probe specializes in a variety of applications. Besides for cardiac, fetal and adult cardiac applications, this probe also produces images for OB-GYN, Abdominal and Transcranial images. It is compatible with a wide array of GE systems including the Logiq P9, Vivid S5, Logiq e, Vivid i and Vivid S70.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Persistent dark bands or a missing sector that do not move with the anatomy.
Root cause: Failed piezoelectric elements or (fetal) a degraded crystal.
Symptoms: Image / signal drops or is noisy when the cable or strain relief is moved.
Root cause: Broken conductors in the transducer cable or at the connector.
Symptoms: Haze, artefacts, reduced penetration, a visible cut or bubble, cracked housing.
Root cause: Lens delamination or a puncture; drop damage.
Symptoms: System does not detect the probe or shows the wrong probe; bent or corroded pins.
Root cause: Damaged connector pins or a wet / dirty connector.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Manufacturer / authorised repair only.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transducer cable Manufacturer / authorised repair only. | Periodic replacement | $$$ | Difficult |
| System connector assembly | Major service part | $$$ | Difficult |
| Acoustic lens / membrane | Periodic replacement | $$$ | Expert-level |
| Strain-relief boot | Periodic replacement | $ | Moderate |
| Needle-guide bracket | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio