FUTURE HEALTH CONCEPTS, INC. · Model FHCSWC36-TL-MB
Overall size 30”W x 26.5”D x 6 x 36”T. Chamber I.D. 26”W x 23”D x 25”T. 8.63 Cubic ft. storage. Digital control. 2 Adjustable shelves. Stainless steel door. Audible and visual over-temp alarm. Accuracy +/- 2 degrees F. Electrical: 120V, 7.5 Amp, 50/60 Hz. .35kw/hr, 1195 btu/hr. UL/CUL listed. Mobile base
Intended use: A mains electricity (AC-powered) device designed to warm blankets with hot dry air. It is typically a high capacity (e.g., 0.2 to 0.5 m3) unit with appropriate perforated shelves for blanket positioning, avoiding contact with very hot surfaces to prevent scorching. Hot dry air is usually circulated to heat blankets in a temperature range from 30 to 90? C (90 to 200? F).
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, overheat alarm.
Root cause: Element or thermal fuse failure.
Symptoms: Setpoint error, alarm.
Root cause: Sensor aging.
Symptoms: Uneven warming, hotspot.
Root cause: Blower wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R.
Consumable per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heating element / thermal fuse Common R&R. | Periodic replacement | $$$ | Moderate |
| Blower fan | Periodic replacement | $$ | Fairly easy |
| Temperature probe Consumable per patient. | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →