Fujifilm · Model FDR D-EVO P-Series
Digital Detectors FDR D-EVO panels are light, portable detectors that can quickly turn your x-ray room into a DR room with incredible versatility, allowing you to perform table and free cassette-based exams with ease. There’s also no need to re-work your existing equipment because FDR D-EVO fits existing standard Buckys and holders. FDR D-EVO P-series panels feature 150 micron pixel pitch and a wide 16-bit dynamic range, as well as Fujifilm’s exclusive, patented ISS (Irradiated Side Sampling) which provides ultra-sharp imaging with advanced dose efficiency. This exclusive design captures the x-ray image from the incident/patient side of the detector, in contrast to traditional designs which capture the image below the scintillator layer. The result is a reduction in the distance for light to spread, allowing the image to be captured where its sharper and stronger as it was always intended. Fujifilm is dedicated both to improving image quality and advancing dose lowering technologies and safe healthcare standards for patients. FDR D-EVO is a perfect example. It features engineering innovations which achieve dose efficiency improvements by as much as 30% compared to Fujifilm CR and as much as 20% over comparable DR detector designs, all while delivering exceptional images.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Panel not detected, images not received, wireless / tethered link drops mid-exam.
Root cause: Detector battery (wireless), Wi-Fi / tether cable, a docking / connector fault, or detector electronics.
Symptoms: Fixed-pattern artefacts, non-uniformity, dead pixels / lines, calibration overdue or failing.
Root cause: Detector gain / offset calibration, a bad pixel map, or physical detector (drop) damage.
Symptoms: Exposure inhibited, kV / mAs error, no exposure captured, AEC over/under.
Root cause: HV generator, tube, filament supply, or an AEC / interface fault.
Symptoms: Tube column, wall stand or table bucky will not track, detent, or auto-position.
Root cause: Drive motor, brake, encoder, or the auto-positioning interface.
Symptoms: Images will not display, process or send; worklist missing; slow processing.
Root cause: Workstation PC, software, or the DICOM / network configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Flat-panel detector (exchange) | Major service part | $$$$ | Moderate |
| Detector battery (wireless) | Periodic replacement | $$ | Easy |
| Tether cable / access point | Periodic replacement | $$ | Fairly easy |
| Positioner drive / brake / encoder | Major service part | $$$ | Difficult |
| Acquisition workstation PC | Major service part | $$$ | Moderate |
| AEC ion chamber | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.