Ethicon · Model G110 Harmonic Scalpel
Frequency: 55.5 kHz (output). Power Requirement: 115 VAC, 60 Hz, 2 Amps, 230 Watts max. Fusing: Two, 3 AG Slow Blow 1⁄4” x 1 1⁄4”, 2 Amps, located in the power entry module. Three, internal, 5 mm x 20 mm Slow Blow, 1@ 1 Amp, 1 @ 0.25 Amps, and 1 @ 1.6 Amps. Air Pressure: 6 psi Max. Air Flow: Approximately 20 liters per minute. Ventilation: At least 2” in rear of the unit and the bottom vents should not be obstructed.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No cut / coag effect, "high impedance" or output-error message, output shuts down.
Root cause: Poor electrode contact or a dry field, a damaged handpiece / cable, a return-electrode (REM/NESSY) fault, or an output-stage fault.
Symptoms: REM / NESSY / plate alarm, monopolar output inhibited.
Root cause: Poorly applied or dried split dispersive pad, a pad-cable fault, or the contact-quality-monitor circuit.
Symptoms: Bipolar or sealing output weak, seal cycle will not complete, instrument not recognised.
Root cause: Bipolar cable / instrument fault or the sealing output stage.
Symptoms: Output power or the REM circuit outside tolerance on the analyser.
Root cause: Output board or REM circuit needs recalibration - anti-static handling required.
Symptoms: Output will not activate from the pedal or the handpiece.
Root cause: Footswitch or handswitch / cable fault.
Symptoms: Unresponsive touchscreen, will not power on or boot, fan / thermal alarm.
Root cause: Display, power supply, control board, or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibration required after replacement; anti-static handling.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable instrument / handpiece cables | Periodic replacement | $$ | Easy |
| Split dispersive (return) pads | Consumable | $ | Easy |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| Bipolar / sealing cable | Periodic replacement | $$ | Easy |
| RF output board Recalibration required after replacement; anti-static handling. | Major service part | $$$$ | Difficult |
| Cooling fan | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns
The devices may have been assembled with an internal component that may cause continuous or inadvertent activation of the device.
Complaints of holes in Tyvek which can compromise the sterile barrier provided by the packaging.
Ethicon Endo-Surgery initiated a voluntary recall due to some lot numbers of the HARMONIC Hand Piece (HP054) and the HARMONIC Blue Hand Piece (HPBLUE), which impacts their compatibility with the HARMONIC Generator 300 (GEN04). The problem occurs after successful set up and initial use of HARMONIC Hand Pieces (HP054 and HPBLUE) with the Generator 300 (GEN04). The procedure is interrupted when