Dräger · Model Polaris
Simply good light! Polaris medical lighting systems deliver cool, efficient and powerful illumination for almost any situation. Highly efficient, state of the art and designed for compatibility, these lights can be upgraded to our future developments and were engineered to protect your investment. The Polaris is simple to handle and operate, easy to clean, safe and reliable. [b] POLARIS 500[/b] The Polaris 500 is a good choice to expand the performance of the surgical light system Polaris. Featuring a total of 60 LEDs the light head provides you with additional illumination capacity. The Polaris 500 can be used as single minor surgical light as well as in several combinations. [b]POLARIS 700[/b] The Polaris 700 features 54 precision reflectors, each holding two LEDs, giving you a total of 108 LEDs. These are bundled together in 6 separate arrays that move in unison when the central handle is rotated. This reflector array design provides adequate shadow control and depth illumination.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: One or more heads / LED segments out, dim head, uneven or discoloured light field.
Root cause: Halogen lamp at end of life or a failed LED module / driver, a lamp-changeover fault, or a power-supply fault.
Symptoms: Light head drifts up or down, will not stay positioned.
Root cause: Spring-arm counterbalance out of adjustment or worn friction brakes.
Symptoms: Brightness or spot size will not adjust from the wall panel or sterile handle.
Root cause: Control PCB, the wall-plate keypad, or the handle switch.
Symptoms: Handle loose or not detected.
Root cause: Worn handle mount / contacts.
Symptoms: Whole head dark, flicker, or will not power on.
Root cause: Failed lamp / LED power supply.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Light-head lamp / LED module | Frequent wear item | $$ | Fairly easy |
| Lamp / LED power supply | Major service part | $$ | Moderate |
| Spring-arm counterbalance / friction-brake kit | Major service part | $$$ | Difficult |
| Control PCB / wall-plate keypad | Major service part | $$ | Moderate |
| Sterilizable handle | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.