DePuy Synthes · Model Mitek Vapr 3
Protection of arthroscopic equipment from damage caused by proximity of radiofrequency energy. Optimized and safe performance for various electrode types without the need to manually adjust the settings. Flexibility and enhanced safety by adjusted ablation speed and depth of tissue effect based on anatomical area and individual patient factors.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No cut / coag effect, "high impedance" or output-error message, output shuts down.
Root cause: Poor electrode contact or a dry field, a damaged handpiece / cable, a return-electrode (REM/NESSY) fault, or an output-stage fault.
Symptoms: REM / NESSY / plate alarm, monopolar output inhibited.
Root cause: Poorly applied or dried split dispersive pad, a pad-cable fault, or the contact-quality-monitor circuit.
Symptoms: Bipolar or sealing output weak, seal cycle will not complete, instrument not recognised.
Root cause: Bipolar cable / instrument fault or the sealing output stage.
Symptoms: Output power or the REM circuit outside tolerance on the analyser.
Root cause: Output board or REM circuit needs recalibration - anti-static handling required.
Symptoms: Output will not activate from the pedal or the handpiece.
Root cause: Footswitch or handswitch / cable fault.
Symptoms: Unresponsive touchscreen, will not power on or boot, fan / thermal alarm.
Root cause: Display, power supply, control board, or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibration required after replacement; anti-static handling.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable instrument / handpiece cables | Periodic replacement | $$ | Easy |
| Split dispersive (return) pads | Consumable | $ | Easy |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| Bipolar / sealing cable | Periodic replacement | $$ | Easy |
| RF output board Recalibration required after replacement; anti-static handling. | Major service part | $$$$ | Difficult |
| Cooling fan | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
The external sterile packaging was found adhered to the internal sterile packaging, potentially compromising sterility.
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Fifty (50) individual units from Lot D20111130 were distributed with the incorrect part number and size etched on the product.
The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Products were made outside of Quality System Regulation, and potentially outside of premarket submission (510k/PMA) for certain devices. The safety or effectiveness of these devices cannot be assured.
Mitek identified that on VAPR Tripolar 90 Degree Suction Electrode the ablation and coagulation buttons are colored incorrectly. The wiring for each button function was confirmed as correct and the non-conformance impacted button color only
Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
Labeled incorrectly as a Blade Up configuration instead of a Blade Down configuration.
Specific units of the DePuy Mitek Sterile Reamers are not labeled correctly, 7 units were labeled as 11MM while the box actually contained a 10 MM reamer
Product may have wrong large product identification label with symbols on the box indicating a S90 VAPR Electrode instead of the VAPR 3.5 hook electrode
Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair
Incorrect suture configuration
Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.
Metal debris in the shaft component of the device
Product sterility may be compromised due to breach in sterile packaging
Product does not meet the required packaging specifications for pouch seal integrity which could lead to a breach in sterility
The inserter shafts bend, as they were manufactured with a different stainless steel than specified.
The device may contain one strand of suture instead of two strands
Mislabeled: The kit contained the incorrect Ligament Passer. The package contained a 9.5 mm instead of the labeled 11.5mm ligament passer