CORPAK Medsystems · Model Polar Enteral Feeding Bags & Set
• Unlike PVC, two-ply EVA/Nylon Polar Bags won’t leach plasticizers into enteral formula. Safe for the environment as well. • Inside of cap is recessed to minimize touch contamination. • Connector strap keeps cap away from funnel during filling to prevent touch contamination. • Pump tubing segment is securely glued to the connector to prevent tubes from blowing off when irrigating or adding meds. • Anti-IV stepped connectors. • Two-ply EVA/Nylon construction eliminates pin holes, adds strength. • EVA/Nylon bags are stronger than PVC, won’t stretch like PVC. • Safety-lock cap provides an airtight seal. • Widest opening of any bag makes spill-free filling easier. • Bag funnel doubles as a handle. • Crystal clear EVA plastic allows you to see meniscus and condition of formula. • Hanger centers the bag and eliminates sagging. • Cap is easy to open and close, even by home patients
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
Symptoms: Set not recognized, door-open error.
Root cause: Dirty or failed door microswitch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R on high-mileage pumps.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic roller / pump mechanism Common R&R on high-mileage pumps. | Periodic replacement | $$$ | Difficult |
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →