
Control-X Medical · Model Chiro System
Convenient control - All tubestand controls are centrally located at the x-ray tube, and the wallstand controls at the operating side of the bucky/grid cabinet. This is for easy fingertip access and operator convenience. Electromagnetic lock system - Electromagnetic brakes provide reliable and precise positioning for both the wallstand and the tubestand. Easy-read tube settings - X-ray tube position settings are in cms/inches and degrees. Dual rope counterbalancing system - It ensures maximum stability and security with minimum effort of movement for both the tubestand and wallstand. Excellent maneuverability - Both the tubestand and wallstand are designed with easy-glide effortless movement for vertical and horizontal (tubestand only) travels. Exceptional versatility - The extended vertical movement of the tube and the wallstand supports examinations on patients with various anatomical characteristics. Model diversity - The wallstand accepts a 17" or 36" bucky or grid cabinet with x-ray grid, and also provides room for an optional AEC chamber. Low dosage exposures - The high frequency techniques of the ConRad generator assure accurate exposure parameters.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, generator error / fault code, images consistently under- or over-exposed.
Root cause: HV generator / inverter fault, an exposure-switch / contactor fault, an HT-cable fault, or a tube-load / thermal interlock.
Symptoms: Output does not match the console, techniques drift, generator calibration fails.
Root cause: Inverter / HV-tank fault, a filament-supply fault, or a calibration issue.
Symptoms: Auto-exposure images consistently over- or under-exposed, chambers not tracking.
Root cause: AEC ion chamber(s) or the AEC calibration.
Symptoms: Column, wall stand, table or bucky will not track / detent; collimator light out; grid errors.
Root cause: Drive / brake / detent hardware, the collimator lamp / drive, or a bucky-motor fault.
Symptoms: Exposures inhibited after a series, tube-heat alarm.
Root cause: Tube housing heat limit reached or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| HV generator / inverter components | Major service part | $$$$ | Expert-level |
| HT cables | Major service part | $$$ | Difficult |
| X-ray tube assembly | Major service part | $$$$ | Expert-level |
| AEC ion chamber | Major service part | $$$ | Difficult |
| Collimator lamp / drive assembly | Periodic replacement | $$ | Moderate |
| Bucky / grid motor | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio