Conmed · Model System 1200
Boost Mode (maximum suction 34 cfm) for smokier condition Normal Mode (maximum suction 20 cfm) for standard conditions. Coordinated activation with the electrosurgical generator, effectively clearing the field of view. The system incorporates a 3-second delay in shut-off to evacuate smoke. Suction Control and Suction Level Indicator, offering the O.R. staff optimal control of smoke evacuation. SmartFilter® Filter Life Indicator, providing visual confirmation when the filter should be changed, with a filter efficiency rating of 99.999% ULPA grade. The filter is completely enclosed to protect the healthcare professional.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Smoke lingers in the field, weak suction at the pencil / wand, capture alarm.
Root cause: Saturated ULPA / charcoal filter, a kinked or blocked hose, or a worn blower.
Symptoms: Filter-change warning, reduced airflow.
Root cause: Filter at end of life or not fully seated.
Symptoms: Rising noise, vibration, airflow falling off over time.
Root cause: Worn blower / turbine bearing.
Symptoms: Evacuator does not start with the ESU or footswitch.
Root cause: Activation sensor or footswitch fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| ULPA + charcoal filter Consumable. | Frequent wear item | $$ | Easy |
| Blower / turbine unit | Major service part | $$$ | Difficult |
| Suction tubing / wand | Frequent wear item | $ | Easy |
| Footswitch / activation sensor | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio