Symptoms: The CPM carriage does not move, stalls, or moves in only one direction; motor noise.
Root cause: Worn drive motor / leadscrew, a hand-control fault, or a control-board fault.
Symptoms: Set flexion / extension angles not reached, or the carriage overshoots / stalls short.
Root cause: Position-sensor / potentiometer drift or a limit-switch fault.
Symptoms: Hand control unresponsive, the patient stop / go-back button does not work.
Root cause: Worn hand control or a damaged cable.
Symptoms: Rough or noisy travel, play in the carriage.
Root cause: Worn carriage bearings / slides.
Symptoms: Will not power on, or short battery run time (if battery-equipped).
Root cause: Power supply / fuse or an aged battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Drive motor / leadscrew | Major service part | $$$ | Difficult |
| Position sensor / potentiometer | Major service part | $$ | Moderate |
| Hand control | Periodic replacement | $$ | Fairly easy |
| Carriage bearings / slides | Major service part | $$ | Moderate |
| Power supply / fuse | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.