Cesca Therapeutics · Model Thermogenesis MT204
The Thermogenesis Thermoline™ plasma thawers represent the latest in rapid and reliable thawing of plasma and cryoprecipitate. There is no contact between the plasma pack and the water There as an agitation action included to further speed up thawing.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "LOW FLD" on the temperature/error display; the thawer is disabled at rest or at the end of a cycle until refilled and RESET is pressed.
Root cause: Water below the recommended operating level, a ground problem, or a corroded level probe.
Symptoms: Unit disabled with an overtemp condition.
Root cause: Incorrectly calibrated operating temperature, a temperature-sensor fault, or a heater/control fault.
Symptoms: No agitation action during a thaw, slow thawing.
Root cause: Agitation motor or drive-linkage fault.
Symptoms: Cycles run long, plasma not fully thawed.
Root cause: Weak heater or poor water circulation.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Corrosion causes false LOW FLD alarms.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water-level probe Corrosion causes false LOW FLD alarms. | Periodic replacement | $$ | Moderate |
| Temperature sensor | Major service part | $$ | Moderate |
| Heater element | Major service part | $$ | Difficult |
| Agitation motor / drive | Major service part | $$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.