Breas Medical AB · Model US
Ventilator Non-continuous (Respirator)
Intended use: A portable, mains electricity (AC-powered) device, which may include rechargeable batteries, intended to assist noninvasive ventilation (i.e., without use of an artificial airway) using continuous positive airway pressure (CPAP) during spontaneous respiration, primarily to treat adult patients affected by obstructive sleep apnoea (OSA); it may also be intended to treat snoring. It is a small desktop unit with controls, and may include a built-in humidifier; the airway pressure may be automatically adjusted to help provide optimal CPAP through use of a sensor (auto CPAP). The device is intended for use in the home but may also be used in healthcare facilities.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Low exhaled tidal volume, high leak alarm, inability to maintain PEEP.
Root cause: Diaphragm/valve degraded by secretions and cleaning cycles.
Symptoms: Unexpected shutdown, persistent low-battery alarm.
Root cause: Aged internal battery or failed PSU.
Symptoms: Leak alarm, low circuit pressure, volume loss.
Root cause: Disconnected or punctured patient circuit / humidifier.
Symptoms: FiO2 out of range, oxygen alarm.
Root cause: Failed solenoid or proportional valve in blender.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Most-replaced wearable part; stock spares.
Consumable, replace per PM schedule.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Exhalation valve assembly Most-replaced wearable part; stock spares. | Frequent wear item | $$ | Fairly easy |
| Flow / differential pressure sensor | Periodic replacement | $$$ | Moderate |
| Bacterial / HEPA filter Consumable, replace per PM schedule. | Consumable | $ | Easy |
| Compressor / turbine module | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →