BOSTON SCIENTIFIC CORPORATION · Model M004CRBS400H0
A mobile assembly of mains electricity (AC-powered) devices intended to cryogenically ablate areas of the endocardium of a beating heart in the treatment of cardiac arrhythmias. It typically includes a console, monitor, connection cable(s), electrocardiogram (ECG) electrodes, and a cryoablation catheter in which extremely low temperatures are achieved through expansion of a pressurized liquid refrigerant (e.g., nitrous oxide) at the catheter tip; monitoring is achieved via ECG displays and sensors within the catheter.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Balloon or tip temperature not dropping to target, long freeze times, poor lesion.
Root cause: Low refrigerant (N2O) supply, a blocked catheter Joule-Thomson section, moisture / ice in the line, or a console gas-delivery fault.
Symptoms: Console does not recognise the catheter, will not arm a freeze.
Root cause: Catheter umbilical connector wear or a damaged catheter hub.
Symptoms: Vacuum alarm, safety interlock, freeze aborts and the catheter auto-thaws.
Root cause: Vacuum pump, a circuit leak, or a monitoring-sensor fault.
Symptoms: Low gas-pressure alarm, or exhaust-scavenging fault.
Root cause: N2O cylinder low, regulator, or the scavenging connection.
Symptoms: Unresponsive display, will not power on.
Root cause: Console power supply or control board.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| N2O gas cylinder / regulator | Frequent wear item | $$ | Fairly easy |
| In-line gas filter / drier | Frequent wear item | $ | Fairly easy |
| Catheter umbilical / connector assembly | Periodic replacement | $$$ | Moderate |
| Vacuum pump | Periodic replacement | $$$ | Difficult |
| Console control board | Major service part | $$$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.