
BK Medical · Model 8809
Intraoperative vascular. General intraoperative. Musculoskeletal. Small part. Interventional.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Persistent dark vertical bands or a missing sector in the image that do not move with the anatomy.
Root cause: Failed piezoelectric elements in the array.
Symptoms: Image drops, flickers or is noisy when the cable or strain relief is moved.
Root cause: Broken conductors in the transducer cable or at the connector strain relief.
Symptoms: Haze, comet artefacts, reduced penetration, visible cut or bubble under the lens.
Root cause: Lens delamination or a puncture letting fluid into the acoustic stack.
Symptoms: Scanner does not detect the probe or shows the wrong probe, bent or corroded connector pins.
Root cause: Damaged scanner-connector pins or a wet/dirty connector.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Repaired by BK Medical or an authorised representative only.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Transducer cable Repaired by BK Medical or an authorised representative only. | Periodic replacement | $$$ | Difficult |
| Scanner connector assembly | Major service part | $$$ | Difficult |
| Acoustic lens / membrane | Periodic replacement | $$$ | Expert-level |
| Strain relief boot | Periodic replacement | $ | Moderate |
| Puncture / needle guide | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.