Syringe drive fluid delivery system for precise control over all fluid flow rates Wash single strips or full plates 96- and 384-well microplates Unique priming trough built into removable microplate carrier for effortless priming and easy maintenance Automatic buffer switching Create, edit, delete and copy up to 75 programs from the keypad Link up to 10 programs to run the most complex routines Multi-user friendly; total programming control for complex research applications, yet easy to use for routine clinical work Programmable shaking duration and intensity Built-in maintenance programs Bottom washing and crosswise aspiration protocols Small footprint with integrated dispense and waste pumps Aerosol cover for safety.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Readings biased or noisy, calibration / self-test fails, poor plate-to-plate reproducibility.
Root cause: Lamp / LED ageing, dirty optics or filters, or a filter-wheel fault.
Symptoms: Plate not drawn in or positioned, carriage-error message, wells read at the wrong position.
Root cause: Worn transport belt / motor or a position-sensor fault.
Symptoms: Wells not aspirated / dispensed evenly, high residual volume, needles blocked.
Root cause: Blocked or bent manifold needles, or a pump / vacuum fault.
Symptoms: A specific wavelength reads wrong, filter not found.
Root cause: Damaged interference filter or a filter-wheel position fault.
Symptoms: Results will not calculate, print or transfer.
Root cause: Onboard software, printer, or the interface.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Lamp / LED source | Periodic replacement | $$ | Fairly easy |
| Interference filters / filter wheel | Major service part | $$ | Moderate |
| Plate-transport belt / motor | Periodic replacement | $$ | Moderate |
| Wash-manifold needles / pump tubing (washers) | Frequent wear item | $ | Fairly easy |
| Optics / detector board | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.