Biomerieux · Model Vitek 2
Three-step setup procedure High level of automation Intuitive, icon-driven software Familiar Windows® format Easy graphical representation of phenotypes One-click result validation.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "Card Transport System Failure" or "Carousel Jam", processing halted.
Root cause: A jammed card, boat, or transport mechanism - resolve the Transport Halted screen; call bioMérieux if it recurs.
Symptoms: "Transmittance Process Failure" or "optical system not available for card type in slot".
Root cause: Contaminated optics - run Cleaning Optics (Cards Processing); call bioMérieux if it recurs.
Symptoms: "Bar Code Read Failure at Slot", or an SCS / barcode-reader conflict.
Root cause: Unreadable card barcode or a reader fault - follow the Bar Code Read Failure procedure or enter codes manually.
Symptoms: "Card Sealer Failure".
Root cause: Sealer fault - call bioMérieux for assistance.
Symptoms: "Card has expired at Slot", or QC / false-resistant results.
Root cause: Expired card, or a card lot subject to a bioMérieux recall (carbapenem/cephalosporin/colistin false-resistant).
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Optical system lamp / detector | Major service part | $$$ | Difficult |
| Card transport mechanism | Major service part | $$$ | Difficult |
| Card sealer assembly (bioMérieux service) | Major service part | $$$ | Expert-level |
| Barcode reader | Major service part | $$ | Fairly easy |
| Test cards | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.
There are 7 reported software anomalies that may affect use of the device.
Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.
bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).
There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card (ref. 421857, lot 8021232403). .
The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card.
Unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits.
False Positive ESBL Phenotype
The VITEK 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) results when using the VITEK 2 FLEXPREP" feature.
False resistant results for Streptococcus anginosus and Streptococcus constellatus strains were reported.
Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).
Invalid calibration with low calibrator S1 while using the product.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Customer reports indicated an increase in the rate of non-detected MRSA in association with the VITEK 2 AST-P631 product.
Missing desiccant.
Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC 700327(TM) strain is positive.
Customers have reported that some VITEK 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.
The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).
The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).
Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.
Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline.
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
The test kits have Out-of-Range-Low (OORL) Flucytosine QC results for Candida krusei ATCC 6258
The outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use.
A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table
There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
The product package did not contain the correct version of the package insert listing the limitations of use and performance requirements.
The firm inventoried and distributed 0.9% Sodium Chloride Saline bags as 0.45% Sodium Chloride Saline bags for use with the VITEK 2 system.
Cards may not load into the incubator correctly causing them to be loaded in the wrong order which could result in the results being linked to the wrong accession number and reporting results erroneously.
The piperacillin/tazobactam drug on the antimicrobial susceptibility cards is providing false susceptible and false resistant results for E. coli, as well as an additional 7 organisms; and the same drug is providing false resistant results for K. pneumoniae.
AMENDED REASON FOR RECALL: The piperacillin/tazobactam drug on the antimicrobial susceptibility cards is providing false susceptible and false resistant results for E. coli, as well as an additional 7 organisms; and the same drug is providing false resistant results for K. pneumoniae.
Foil pouch was punctured by drug susceptibility cards during packaging, allowing moisture into pouch, resulting in inaccurate and/or incorrect drug susceptibility results.
Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
Some of the cards were stamped with an incorrect card code causing the system to read and report the cards incorrectly.
The cards were stamped with an incorrect card code that causes the system to read and analyze the cards incorrectly.