Bio-Rad Laboratories, Inc. · Model NGC 100
The guided fluidics selection tool (patent pending), coupled with the module discovery system, enables a single-click application-based system setup through automated module function recognition. The Tier Rotate™ Feature allows optimal placement of valves and detectors to minimize the flow path. Compact footprint - fits in a refrigerator, lab bench, or coldroom. Point-to-Plumb™ Feature shows the step-by-step plumbing of the system using LEDs. ChromLab™ Software provides powerful system control, automated methods, and analysis options.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: ChromLab connection-failure message, no instrument control, touch screen shows no link.
Root cause: Wrong network name / IP, a firewall blocking ChromLab services, or a network / cabling fault between the PC and the instrument.
Symptoms: Unstable baseline, pressure spikes or a high-pressure trip, poor gradient reproducibility.
Root cause: Worn pump-head seals or check valves, air in the inlet lines, or a partially blocked in-line filter / column.
Symptoms: Incorrect gradient, salt concentration off, valve not switching.
Root cause: Sticking or worn rotary inlet valve, or degraded valve tubing.
Symptoms: Noisy or drifting signal, low lamp intensity warning, poor peak detection.
Root cause: Aged UV lamp, air bubble or contamination in the flow cell, or a fouled conductivity cell.
Symptoms: A stacked module is not detected by the discovery system, single-click setup fails.
Root cause: Module interconnect not seated, or a communication fault on that module.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Routine wear item on a chromatography pump.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump-head seal and check-valve kit Routine wear item on a chromatography pump. | Frequent wear item | $$ | Moderate |
| In-line / inlet filter set | Frequent wear item | $ | Fairly easy |
| Inlet / buffer-blend valve tubing kit | Frequent wear item | $ | Fairly easy |
| UV detector lamp | Periodic replacement | $$$ | Moderate |
| Flow-cell / conductivity-cell assembly | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
It was discovered that the Variant II Turbo Link Calibrator Values card contains incorrect values for IFCC units. The numbers are missing a decimal point.