Bio-Med Devices · Model Bio - Med Devices 2004
The 2004 is a High-Low Flow Blender (with the emphasis on High). It has 3 DISS ports (left, right and bottom). If anything is screwed onto the right side port, the 12lpm bleed flow is activated and the blender's accuracy range is from 2-108 lpm. This bleed flow will be active whenever the blender is connected to an air and oxygen gas source (even if the flow meters are turned off). If nothing is connected to the right side port (the bleed is not active), the accuracy range of the blender is between 15-120lpm.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: FiO2 or flow does not match the setting, mix alarm, hypoxic-guard trips.
Root cause: Blender proportioning valve, a supply-pressure imbalance, or a worn flowmeter / needle valve.
Symptoms: Blender alarm, will not deliver at flow, one gas reads low.
Root cause: Wall / cylinder supply low or one gas missing, or an inlet filter / check-valve fault.
Symptoms: Hiss, high gas consumption, will not hold set pressure.
Root cause: Worn O-rings / seals at inlet or outlet fittings, or the blender bleed.
Symptoms: Float hangs up, flow does not track the knob.
Root cause: Dirty or scratched flow tube, or a worn needle valve.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Proportioning valve / blender core | Major service part | $$$ | Difficult |
| Inlet filters / check valves | Periodic replacement | $ | Fairly easy |
| Flowmeter (flow tube / needle valve) | Major service part | $$ | Moderate |
| O-ring / seal kit | Frequent wear item | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio