Bayer · Model CONTOUR
• No Coding™ technology – you never need to code your meter. • Easy viewing of sample fills. The window makes it easy to see when the strip fills with blood using Sip-in Sampling® technology. • Meter automatically turns on when test strip is inserted and is ready to test. • Flip-top bottle opens, closes and handles easily. • Bottle design keeps strips fresh up to 6 months after opening.
No documents available for this device yet.
Synced live from the FDA openFDA API.
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.
All 10- or 25- count glucose test strip vials, either loose vials or found in patient starter kits of Contour, Contour TS and Contour" USB blood glucose meters are being recalled a packaging issue which, under certain conditions, can cause some test strips packaged in some small-count vials to operate outside of US FDA-cleared performance specifications and produce a low-bias reading during blood
InaccurateTest Strip Results: Results in blood glucose readings with a positive bias are outside the product specifications. Patient test results may demonstrative results 5 percent to 17 percent higher.
Incorrect Unit of Measure displayed : The meters report results in the wrong unit of measurement for Canadian users.