Baxter International · Model Sabratek 6060
Baxter has ended service on this product. Please see one of the available 3rd party companies for service and/or sales. The Baxter Sabratek 6060 is an ambulatory multi-therapy device that can be used for enteral, epidural, subcutaneous, arterial, and intravenous.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: The pump alarms and displays an error message / code and stops.
Root cause: A subsystem fault indexed in the service manual - motor / drive, a sensor, set-detection, or a board.
Symptoms: Repeated occlusion alarms, delivery stops, or an occlusion is missed.
Root cause: Kinked or clamped line, a closed roller clamp, a positional patient tube, pressure-sensor drift, or an improperly loaded set.
Symptoms: Frequent air-in-line alarms, delivery stops.
Root cause: Air in the set, an improperly loaded set, or an ultrasonic air-detector fault.
Symptoms: Delivered volume outside tolerance on the annual test, especially at low rates.
Root cause: Worn pumping mechanism / plunger drive.
Symptoms: Short run time, low-battery early after a full charge.
Root cause: Aged battery pack.
Symptoms: Set will not load, door will not latch, plunger grip poor.
Root cause: Worn door latch, plunger jaws, or anti-free-flow mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Pumping mechanism / rotor / plunger drive | Major service part | $$$ | Difficult |
| Pressure / air-detector sensor assembly | Major service part | $$ | Moderate |
| Door / latch / anti-free-flow assembly | Major service part | $$ | Moderate |
| Drive motor / gearbox | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Failures within the PCA profile as well as incidents resulting in interruptions of therapy in various profiles.
Possible shorting of some printed circuit boards could result in a possible interruption of therapy or cause the pump to fail the Power-On Self Test.
The inlet and outlet lines of the administration set may be reversed, which would result in reverse flow.
A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
The pump may free-flow if the door to the 6060 pump is not closed prior to loading the pump in to the lock box. The pump will alarm 'Door Open', but if the pump is turned off to silence the alarm while remaining in the lock box, the fluid will free-flow.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio