Baxter has ended service on this product. Please see one of the available 3rd party companies for service and/or sales. The Baxter Sabratek 6060 is an ambulatory multi-therapy device that can be used for enteral, epidural, subcutaneous, arterial, and intravenous.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Uncontrolled flow when door open.
Root cause: Worn roller or failed free-flow valve.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
| IV administration set Consumable; per patient. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Failures within the PCA profile as well as incidents resulting in interruptions of therapy in various profiles.
Possible shorting of some printed circuit boards could result in a possible interruption of therapy or cause the pump to fail the Power-On Self Test.
The inlet and outlet lines of the administration set may be reversed, which would result in reverse flow.
A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
The pump may free-flow if the door to the 6060 pump is not closed prior to loading the pump in to the lock box. The pump will alarm 'Door Open', but if the pump is turned off to silence the alarm while remaining in the lock box, the fluid will free-flow.