Bard · Model Criticore
Continuous temperature display eliminates the need for other invasive temperature probes. Automated measurement provides for faster assessment of the patients condition. Reduces supply cost and improves clinician productivity. Reduces patient’s risk of infection through fewer invasive probes and by use of a closed system.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: "---" or implausible temperature, large offset vs a reference, no continuous trend.
Root cause: Failed or disconnected temperature-sensing Foley catheter, a damaged sensor cable, or a monitor input fault.
Symptoms: Output not accumulating, implausible volume, drainage not registering.
Root cause: Blocked or kinked drainage tubing, an air-lock, or a load-cell / optical sensor fault in the monitor.
Symptoms: Intermittent temperature or output signal, connector loose.
Root cause: Repeated flexing and connect/disconnect at the patient cable.
Symptoms: Short battery run time, unit powers off, will not charge.
Root cause: Aged battery or charging-circuit fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-patient consumable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Temperature-sensing Foley catheter Single-patient consumable. | Consumable | $ | Easy |
| Reusable patient / sensor cable | Frequent wear item | $$ | Easy |
| Urine-output measurement sensor / load cell | Periodic replacement | $$$ | Difficult |
| Rechargeable battery | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.