Arjo · Model Sara Combilizer
Supporting mobilization Safe working load of 440lbs Sideways tilting in all positions, for pressure re-distribution and rehab purposes Adjustable shoulder supports, foot plate, arm rests and lumbar support.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No powered movement, hand control dead, will not change position.
Root cause: Discharged or incorrectly inserted battery, an active emergency stop, an activated pinch-protection bar, an obstruction, patient over 200 kg, or a faulty hand control / cable.
Symptoms: Mode selection (stretcher / chair / tilt table) will not switch.
Root cause: Control-panel or configuration fault, or an interlock condition.
Symptoms: One movement will not run, runs one direction only, noisy or slow.
Root cause: Worn linear actuator or a limit-switch fault.
Symptoms: Short run time, frequent low-battery, unit dies mid-transfer.
Root cause: Aged battery pack or charger fault.
Symptoms: Slings fraying or coming apart, sling rope knots slipping.
Root cause: Sling wear or insufficient rope length past the knot (subject of recalls).
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Inspect before every use; retire per Arjo guidance and the recall notices.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery pack | Periodic replacement | $$ | Easy |
| Hand control | Periodic replacement | $$ | Easy |
| Linear actuator (tilt / back / height) | Major service part | $$$ | Difficult |
| Slings / lifting accessories Inspect before every use; retire per Arjo guidance and the recall notices. | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
There is the potential for the slings to come apart, potentially allowing injury to the patient.
The sling ropes were manufactured with insufficient rope length past the knot to prevent the knot from coming loose.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.