Abiomed, Inc. · Model 1000413
Impella CP with SmartAssist (Pump 382) pump set.
Intended use: A sterile tubular device with a built-in electrically-powered axial flow pump (e.g., a micro motor with an impeller) at the distal end, designed to provide circulatory assistance to the heart by pumping blood during heart failure. It is inserted through the lumen of the aorta and into the left ventricle, surgically or percutaneously via an artery, and connected to an external control unit; it moves blood from the left ventricle to the aorta across the aortic valve. It is typically used emergently post acute myocardial infarction (AMI), electively for high-risk cardiac catheterization, for open chest surgery, and during transport between hospitals. This is a single-use device.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Flow drops below the set level, flow alarms, haemodynamic support falling.
Root cause: Pump position shifted (too far in / out or across the valve), suction from low LV volume, a kinked catheter, or a motor / rotor fault.
Symptoms: Purge pressure high or low, purge-flow alarm, purge cassette fault.
Root cause: Purge cassette / tubing occlusion or air, wrong purge-fluid concentration, or a purge-pump fault in the controller.
Symptoms: Damped or absent placement (aortic) pressure waveform, "Impella position unknown".
Root cause: Optical / pressure sensor in the catheter, air in the pressure lumen, or a catheter-connector fault.
Symptoms: Automated Impella Controller alarms, will not run on battery, screen or interface fault.
Root cause: Controller hardware fault or an aged controller battery.
Symptoms: Intermittent motor current or signal, connection alarms.
Root cause: Worn catheter connector or the controller connector cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-use consumable.
Single-use; the pump/rotor is part of the catheter.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Automated Impella Controller (AIC) battery | Periodic replacement | $$$ | Fairly easy |
| Purge cassette / tubing set Single-use consumable. | Consumable | $$ | Easy |
| Connector / mobile-cable assembly | Periodic replacement | $$$ | Easy |
| Impella catheter / pump set Single-use; the pump/rotor is part of the catheter. | Consumable | $$$$ | Easy |
| AIC console (loaner / exchange) | Major service part | $$$$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.