Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 15, 2025
Brand Name: Broselow
Product Name: Broselow Domestic Complete ALS Organizer
Model/Catalog Number: 7730ALS
Software Version: Not Applicable
Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.
Component: No
Reason: Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
Root cause: Device Design
Recalling firm: SunMed Holdings, LLCType: Emergency Response Safety KitDistribution: Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Singapore.Recall #Z-1116-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 15, 2025
Brand Name: Broselow
Product Name: BROSELOW" FILLED BROSELOW ORGANIZER
Model/Catalog Number: 7730IALS
Software Version: Not Applicable
Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.
Component: no
Reason: Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
Root cause: Device Design
Recalling firm: SunMed Holdings, LLCType: Emergency Response Safety KitDistribution: Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Singapore.Recall #Z-1117-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as: NERVE BLOCK TRAY 2, Kit SKU DYNJRA2713
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Spinal Epidural Anesthesia KitDistribution: US Nationwide distribution.Recall #Z-1186-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) CATH LAB PACEMAKER PACK-LF, Kit SKU DYNJ64837C;
2) LOOP RECOVERY PACK, Kit SKU DYNJT7034;
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Cardiac Catheterization KitDistribution: US Nationwide distribution.Recall #Z-1187-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) PACER / CATH LAB PACK-LF, Kit SKU DYNJ0972604A;
2) PACEMAKER/ICD PACK, Kit SKU DYNJ33822L;
3) VENOUS ACCESS PACK, Kit SKU DYNJ45640B;
4) PACEMAKER SET UP, Kit SKU DYNJ63377D;
5) PACK PACEMAKER CUSTOM,9909178, Kit SKU DYNJ65011D;
6) HEART BASIC OPEN PACK, Kit SKU DYNJ69143C;
7) SCC VEIN PACK, Kit SKU DYNJ69313;
8) CV PACK, Kit SKU DYNJ86007B.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Cardiovascular Procedure KitDistribution: US Nationwide distribution.Recall #Z-1188-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) BREAST-HERNIA-PORT CDS-LF, Kit SKU CDS984853L;
2) BREAST BIOPSY PACK, Kit SKU DYNJ45641B;
3) DERMATOLOGY PACK, Kit SKU DYNJT3658A.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Ct Biopsy TrayDistribution: US Nationwide distribution.Recall #Z-1189-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) MIDDLE EAR CDS, Kit SKU CDS984347O;
2) HEAD AND NECK CDS-LF, Kit SKU CDS984857N;
3) NECK PACK-LF, Kit SKU DYNJ0966130K;
4) NECKLINE PACK-LF, Kit SKU DYNJ51440B;
5) THYROID PACK, Kit SKU DYNJ58742J;
6) ENT I, Kit SKU DYNJ63781B;
7) NECK PACK, Kit SKU DYNJ82751B;
8) NOSE PACK, Kit SKU DYNJ82753C;
9) WMC COCHLEAR IMPLANT ADD ON, Kit SKU DYNJ907206G.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Ear, Nose, And Throat Surgical TrayDistribution: US Nationwide distribution.Recall #Z-1190-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) SMJ LAP CHOLE PACK-LF, Kit SKU DYNJ19974P;
2) DEPAUL ABDOMINAL ENDOSCOPY-LF, Kit SKU DYNJ20027K;
3) LAPAROSCOPY PACK-LF, Kit SKU DYNJ20468J;
4) SMJ LAPAROSCOPY PACK, Kit SKU DYNJ25763N;
5) SMJ LAPAROSCOPY PACK, Kit SKU DYNJ25763O;
6) BWNBORO LAP CHOLE PACK-LF, Kit SKU DYNJ32836B;
7) MINOR SET UP PACK, Kit SKU DYNJ43272K;
8) PK, GEN LAPAROSCOPIC CHOLE, Kit SKU DYNJ43976C;
9) GV LAP CHOLE PACK, Kit SKU DYNJ46939Q;
10) LAP CHOLE PACK, Kit SKU DYNJ49688C;
11) GENERAL ROBOTIC LAP, Kit SKU DYNJ51347N;
12) LAP CHOLE PACK, Kit SKU DYNJ54081F;
13) PEDIATRIC MOSES, Kit SKU DYNJ58864B;
14) LAPAROTOMY PACK, Kit SKU DYNJ59313B;
15) CRMC LAP CHOLE DIAG LAP-LF, Kit SKU DYNJ61095F;
16) LAP GASTRIC PACK, Kit SKU DYNJ62065A;
17) MINOR PACK, Kit SKU DYNJ64174A;
18) GENERAL SURGERY PACK, Kit SKU DYNJ64258I;
19) GENERAL LAP SHARED-LF, Kit SKU DYNJ64786D;
20) CHOLE PACK, Kit SKU DYNJ67686B;
21) GEN SURG MINOR PACK, Kit SKU DYNJ69142F;
22) LAP CHOLE PACK, Kit SKU DYNJ81695B;
23) GENERAL LAPAROSCOPY, Kit SKU DYNJ900901K;
24) GEN MINOR, Kit SKU DYNJ900906L;
25) ROBOTIC GENERAL, Kit SKU DYNJ906861D;
26) RMC GENERAL LAPAROSCOPIC, Kit SKU DYNJ907087C;
27) LAP CHOLE LMC CDS, Kit SKU DYNJ911116A;
28) MINOR GENERAL PACK, Kit SKU DYNJT3884;
29) GENERAL LAPAROSCOPY PACK, Kit SKU DYNJT4504;
30) MINOR SURGERY PACK-LF, Kit SKU PHS109062D.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Laparoscopy KitDistribution: US Nationwide distribution.Recall #Z-1191-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) MEM MAXILLOFACIAL PACK-LF, Kit SKU DYNJ0660040R;
2) PLASTICS VCH, Kit SKU DYNJ39225K;
3) WT BREAST PACK, Kit SKU DYNJ46620I;
4) ENDO / AUG PACK, Kit SKU DYNJ59250I;
5) KMC PLASTIC/GEN PACK-LF, Kit SKU DYNJ65980J;
6) PLASTICS PACK, Kit SKU DYNJ68288A;
7) IMPLANT REMOVAL KIT, Kit SKU MNS13035;
8) PLASTICS PACK-LF, Kit SKU PHS41746D.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Plastic Surgery And Accessories KitDistribution: US Nationwide distribution.Recall #Z-1192-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) DRAWER 1 AIRWAY, Kit SKU ACC010270A;
2) DRAWER 2, Kit SKU ACC010690;
3) DRAWER 3 SYR/DRS/MON/NG 1 OF 3, Kit SKU ACC011073;
4) HYPOSPADIUS PACK, Kit SKU DYNJ00282Q;
5) SJ NEURO PACK RFID-LF, Kit SKU DYNJ0101360K;
6) CENTRAL LINE PACK-LF, Kit SKU DYNJ0220136S;
7) SPINE PACK-LF, Kit SKU DYNJ0376635Q;
8) ORTHO SPINAL PACK-LF, Kit SKU DYNJ0460768T;
9) W.C. MINOR LAPAROTOMY PK-LF, Kit SKU DYNJ0509186J;
10) NEURO SHUNT PACK-LF, Kit SKU DYNJ0578815W;
11) ANCILLARY SPINE PACK-LF, Kit SKU DYNJ0610290T;
12) MINOR LAPAROTOMY PACK-LF, Kit SKU DYNJ0753436P;
13) KNEE SCOPE PACK-LF, Kit SKU DYNJ0890502R;
14) PEDIATRIC PACK-LF, Kit SKU DYNJ28212L;
15) ANTERIOR CERVICAL PACK, Kit SKU DYNJ31341P;
16) PEDIATRIC MINOR PROCEDURE PACK, Kit SKU DYNJ37615J;
17) LAPAROSCOPY PACK, Kit SKU DYNJ39215N;
18) MAJOR LAP VCH, Kit SKU DYNJ39217P;
19) NEURO PACK, Kit SKU DYNJ39220N;
20) CAS HAND PK, Kit SKU DYNJ39468G;
21) RFT ISC NOBLES MINOR PACK, Kit SKU DYNJ43954B;
22) PED OPEN HEART PACK, Kit SKU DYNJ45373G;
23) MAJOR GENERAL PACK-LF, Kit SKU DYNJ51436F;
24) MINOR PK-LF, Kit SKU DYNJ51437C;
25) VEIN ABLATION & EXCISION ASC, Kit SKU DYNJ52586C;
26) SHOULDER PACK, Kit SKU DYNJ56889B;
27) LAP CHOLE PACK, Kit SKU DYNJ57206F;
28) TRANSGENDER MASTECTOMY 2, Kit SKU DYNJ58257I;
29) SHOULDER ARTH PACK, Kit SKU DYNJ58410D;
30) MAJOR PACK, Kit SKU DYNJ58528J;
31) PACK BREAST, Kit SKU DYNJ60157F;
32) CH TRANSPHENOIDAL PACK, Kit SKU DYNJ61696I;
33) RR-SPINE PACK, Kit SKU DYNJ62643B;
34) MINOR PACK, Kit SKU DYNJ64238D;
35) PBMS NECK PACK-LF, Kit SKU DYNJ64961C;
36) BREAST PACK, Kit SKU DYNJ64988C;
37) MINOR PACK 180110, Kit SKU DYNJ65447C;
38) PACK,V-P SHUNT, Kit SKU DYNJ65472C;
39) BASIC PACK, Kit SKU DYNJ65996D;
40) MINOR SS, Kit SKU DYNJ66094A;
41) CRMC ROBOTIC GENERAL PACK, Kit SKU DYNJ66613D;
42) PEDIATRIC GENERAL PACK, Kit SKU DYNJ66644F;
43) URO/GYN PACK, Kit SKU DYNJ66675F;
44) SHOULDER SCOPE PACK, Kit SKU DYNJ68707B;
45) SPECIALTY FACIAL PLASTICS PACK, Kit SKU DYNJ68844B;
46) SPECIALTY FACIAL PLASTICS PACK, Kit SKU DYNJ68844C;
47) ACF PACK, Kit SKU DYNJ69112C;
48) SM OR-DR. YEH PACK-LF, Kit SKU DYNJ69705C;
49) SHOULDER ARTHROSCOPY PACK, Kit SKU DYNJ81752C;
50) C BIRTH SET UP PACK, Kit SKU DYNJ81994C;
51) TRIAL PACK, Kit SKU DYNJ82770A;
52) PROSTATE ROBOTIC PACK, Kit SKU DYNJ84120C;
53) HAND PACK, Kit SKU DYNJ85033A;
54) ARTHROSCOPY PACK, Kit SKU DYNJ85197A;
55) GU-GYN-LITHO PACK, Kit SKU DYNJ86013A;
56) DBD-ARTHROSCOPY PACK, Kit SKU DYNJ86963A;
57) BREAST PACK, Kit SKU DYNJ87587A;
58) PACK,PEDIATRIC MINOR, Kit SKU DYNJ88718;
59) UNIVERSAL-ASC, Kit SKU DYNJ900071C;
60) BASIC, Kit SKU DYNJ905869G;
61) PACK,PEDIATRIC MAJOR, Kit SKU DYNJ906888F;
62) HAND/FOOT PACK, Kit SKU DYNJT3131A;
63) SHOULDER PACK, Kit SKU DYNJT3266;
64) LUMBAR PACK, Kit SKU DYNJT4178;
65) CVS ICU PACK, Kit SKU PHS41728D;
66) SINGLE TAP BLOCK TRAY, Kit SKU SPEC0156A.
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as: HEAD SHAVE-LATEX SAFE KIT, Kit SKU P903933B
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Suture KitDistribution: US Nationwide distribution.Recall #Z-1194-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880;
2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Dressing Change TrayDistribution: US Nationwide distribution.Recall #Z-1195-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) DEPAUL BASIC NEURO-LF, Kit SKU DYNJ20028L;
2) NEURO MINOR PACK, Kit SKU DYNJ46546M;
3) NEURO SHUNT PACK, Kit SKU DYNJ54396K;
4) NEURO PACK, Kit SKU DYNJ59397Q;
5) NEURO BASIC, Kit SKU DYNJ904052L;
6) PK NEURO SHUNT BUMC, Kit SKU DYNJT6428.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Neurological TrayDistribution: US Nationwide distribution.Recall #Z-1196-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) RR-GYN LAPAROSCOPY PACK-LF, Kit SKU DYNJ0395022O;
2) PELVISCOPY PACK-LF, Kit SKU DYNJ0415776Q;
3) DIVA PACK, Kit SKU DYNJ04811O;
4) DA VINCI PROSTATE/COLPOPEXY, Kit SKU DYNJ44864S;
5) MH ROBOTICS PACK, Kit SKU DYNJ50880L;
6) DELNOR LAPAROSCOPIC GYN PACK, Kit SKU DYNJ82244B;
7) ROBOT PACK, Kit SKU DYNJ82761B;
8) GENERAL ROBOTICS PACK, Kit SKU DYNJT4964.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Gynecological Laparoscopic KitDistribution: US Nationwide distribution.Recall #Z-1197-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as: CIRCUMCISION PACK, Kit SKU DYNJ16826O
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Circumcision TrayDistribution: US Nationwide distribution.Recall #Z-1198-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) C-SECTION PACK, Kit SKU DYNJ35494C;
2) SMJ C-SECTION PACK-LF, Kit SKU DYNJ38582O;
3) RR-L&D C-SECTION PACK-LF, Kit SKU DYNJ82084F.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Cesarean Section TrayDistribution: US Nationwide distribution.Recall #Z-1199-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 12, 2025
Medline medical procedure convenience kits labeled as:
1) WATKINS SPINE PACK-LF, Kit SKU DYNJ0275635M;
2) SM-SPINE PACK-LF, Kit SKU DYNJ0395037X;
3) TOTAL HIP PACK, Kit SKU DYNJ04809N;
4) MAIN/LAMINECTOMY PACK-LF, Kit SKU DYNJ0565499AN;
5) SCC/SPOC HAND & FOOT PACK, Kit SKU DYNJ27164Q;
6) ORTHO SPINE, Kit SKU DYNJ40221G;
7) URO GYN PACK, Kit SKU DYNJ45311J;
8) LAMINECTOMY PACK, Kit SKU DYNJ46330C;
9) DJ ORTHO SPINE, Kit SKU DYNJ48980M;
10) CERVICAL PACK, Kit SKU DYNJ52995F;
11) NORTH BACK FUSION PACK-LF, Kit SKU DYNJ56483L;
12) KNEE ARTHROSCOPY CHRISTUS, Kit SKU DYNJ61200B;
13) KNEE ARTHROSCOPY PACK-LF, Kit SKU DYNJ66150;
14) HIP PACK, Kit SKU DYNJ66630;
15) OHNS FREE FLAP A PACK, Kit SKU DYNJ69065F;
16) LUMBAR PACK, Kit SKU DYNJ69113C;
17) LAMINECTOMY PACK, Kit SKU DYNJ69401D;
18) LAMINECTOMY PACK, Kit SKU DYNJ81013D;
19) HIP/SHOULDER ARTHROSCOPY PACK, Kit SKU DYNJ84275A;
20) FLASC ACL PACK, Kit SKU DYNJ88971;
21) SPINE PACK MERCY GALENA, Kit SKU DYNJ89943;
22) SPINAL FUSION MT CARMEL, Kit SKU DYNJ900173J;
23) HAND, Kit SKU DYNJ904945G;
24) LAMINECTOMY, Kit SKU DYNJ909505D;
25) SHOULDER SPLIT, Kit SKU DYNJ910543F;
26) SHOULDER SPLIT, Kit SKU DYNJ910543G;
27) CRANIOTOMY PACK, Kit SKU DYNJT4724;
28) SPINE PACK, Kit SKU DYNJT5184.
Reason: Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Root cause: Nonconforming Material/Component
Recalling firm: Medline Industries, LPType: Orthopedic TrayDistribution: US Nationwide distribution.Recall #Z-1200-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 24, 2025
Brand Name: Zimmer Tourniquet Systems
Product Name: A.T.S. 5000TS Tourniquet Systems
Model/Catalog Number: 60500010100
Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S. 5000TS
Product Description: A.T.S 5000TS Tourniquet Systems
Component: N/A
Reason: Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Root cause: Under Investigation by firm
Recalling firm: Zimmer Surgical IncType: Tourniquet, PneumaticDistribution: International distribution to the countries of Canada and EMEA only.Recall #Z-1201-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 24, 2025
Brand Name: Zimmer Tourniquet Systems
Product Name: A.T.S. 3200TS Tourniquet Systems
Model/Catalog Number: 60320010100
Software Version: Software version v2.06 or prior are within scope of the recall for A.T.S. 3200TS
Product Description: A.T.S 3200TS Tourniquet Systems
Component: N/A
Reason: Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.
Root cause: Under Investigation by firm
Recalling firm: Zimmer Surgical IncType: Tourniquet, PneumaticDistribution: International distribution to the countries of Canada and EMEA only.Recall #Z-1202-2026
Open, ClassifiedPosted Jan 28, 2026·Initiated Dec 26, 2025
Chest Drainage Units and Accessories:
Product Code Description
1. 8888571562 Cardinal Health Sentinel Seal CDU
2. 8888571513 Cardinal Health Sentinel Seal CDU, Dual Drain
3. 8888571489 Cardinal Health Sentinel Seal CDU, with Easy Change
Connector
4. 8888571370 Cardinal Health Altitude CDU, Dry Suction, with Easy Change Connector
5. 8888571299 Cardinal Health Aqua-Seal CDU, Wet Suction
6. 8888571406 Cardinal Health Aqua-Seal CDU, Wet Suction, Dual Drain
7. 8888571315 Cardinal Health Aqua-Seal CDU, Wet Suction, with Easy
Change Connector
8. 8884713100 Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000
mL
9. 1814713105 OUSOnly - Cardinal Health Thora-Seal Basic CDU, One Chamber, 2000 mL (Bottle only)
10. 1180571570 OUSOnly - Cardinal Health Thora-Seal CDU, Two Chamber System, 2600 mL
11. 8884713308 Cardinal Health Thora-Seal CDU, Three Chamber System, 2500 mL
12. 8884713900 Cardinal Health Thora-Seal Replacement Collection Chamber,
2500 mL
Reason: The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Root cause: Error in labeling
Recalling firm: Cardinal Health 200, LLCType: Apparatus, Suction, Operating-Room, Wall Vacuum PoweredDistribution: Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico.Recall #Z-1203-2026