Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 30, 2026
Radial Jaw 4 Jumbo Biopsy Forceps with the following descriptions and UPN:
Radial Jaw 4 Jumbo 240cm Box 5; Outer Box UPN: M00513363; Inner Box UPN: M00513360.
Reason: Potential for breach in sterile barrier of device packaging.
Root cause: Process control
Recalling firm: Boston Scientific CorporationType: Forceps, Biopsy, Non-ElectricDistribution: US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN,
KY, MA, MD, ME, MN, MO, MS, MT, NC, NJ, NM, NY, OH, PA, PR, SC, TN, TX, UT, VA, VT, WA, WI and International distribution pending.Recall #Z-2840-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Quantum 4Front 2
Model/Catalog Number: Quantum 4Front 2
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2841-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Quantum Outback
Model/Catalog Number: Quantum Outback
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2842-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Q6 Edge 3 Stretto
Model/Catalog Number: Q6 Edge 3 Stretto
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2843-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: R-TRAK
Model/Catalog Number: R-TRAK
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2844-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Q6 Edge HD
Model/Catalog Number: Q6 Edge HD
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2845-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum rehab
Product Name: Q6 Edge 2.0
Model/Catalog Number: Q6 Edge 2.0
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2846-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Quantum 1450
Model/Catalog Number: Quantum 1450
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2847-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Q6 Edge 2.0 (ilevel)
Model/Catalog Number: Q6 Edge 2.0 (ilevel)
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2848-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Q6 Edge 3
Model/Catalog Number: Q6 Edge 3
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2849-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Q6 Edge 4
Model/Catalog Number: Q6 Edge 4
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2850-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Quantum 4Front 2HD
Model/Catalog Number: Quantum 4Front 2HD
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2851-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: 4Front
Model/Catalog Number: 4Front
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2852-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 25, 2026
Brand Name: Quantum Rehab
Product Name: Quantum Rival
Model/Catalog Number: Quantum Rival
Software Version: 2.0.11, new version: 2.0.13
Product Description: Wheelchair, Powered (ITI), 890.3860
Component: Controller
Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root cause: Under Investigation by firm
Recalling firm: Pride Mobility Products Corp.Type: Wheelchair, PoweredDistribution: Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN,
AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH,
SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.Recall #Z-2853-2026 Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 16, 2026
Brand Name: WASPLab
Product Name: WASPLab
Model/Catalog Number: W087
Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2
Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.
Reason: Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.
Root cause: Under Investigation by firm
Recalling firm: Copan WASPType: Calculator/Data Processing Module, For Clinical UseDistribution: US Nationwide distribution in the state of Arizona.Recall #Z-2854-2026
Open, ClassifiedPosted Jul 31, 2026·Initiated Mar 23, 2026
Nerivio Infinity. Model/Catalog Number: FGD000100.
Software Version: FW Version 4.1.2.0, App version 7.28.0.11
Reason: Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.
Root cause: Component design/selection
Recalling firm: THERANICA BIOELECTRONICS LTDType: Distal Transcutaneous Electrical Stimulator For Treatment Of Acute MigraineDistribution: The product's entry point to the US is a 3PL warehouse from which it is being shipped to DME distributors or specialty pharmacy who perform the shipment to the end-userRecall #Z-2855-2026
Open, ClassifiedPosted Jul 31, 2026·Initiated Jun 19, 2026
munevo DRIVE and munevo DRIVE r-net
Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
Reason: A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
Root cause: Software Design Change
Recalling firm: MUNEVO GMBHType: Components, WheelchairDistribution: Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.Recall #Z-2856-2026
Open, ClassifiedPosted Jul 30, 2026·Initiated May 11, 2026
Power Over Fiber Base Module
Reason: Device not certified - no laser product report was filed with CDRH prior to distribution.
Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device.
Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56.
Missing certification and proper manufacturer's ID label.
Was not accompanied by compliant User Information.
Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards.
Sales information did not convey required class designation and warning statement
Root cause: Radiation Control for Health and Safety Act
Recalling firm: Terex UtilitiesType: Fiber Optic Communication And Data TransferDistribution: U.S. Nationwide Distribution and International DistributionRecall #Z-2745-2026
Open, ClassifiedPosted Jul 30, 2026·Initiated May 13, 2026
Pack Angiography NCMC 3/Cs BANNER, REF: NCAG62I, containing Medline Namic Angiographic Star Off Handle Manifolds
Reason: Cardiovascular procedure kits contain manifolds, the subject of another recall, that may have particulate within the fluid path, which has the potential to be introduced into blood circulation and become lodged within blood vessels.
Root cause: Material/Component Contamination
Recalling firm: Medical Action Industries, Inc. 306Type: Cardiovascular Procedure KitDistribution: US: AZ, CORecall #Z-2792-2026
Open, ClassifiedPosted Jul 29, 2026·Initiated May 8, 2026
Brand Name: GEM Premier 5000
Product Name: GEM Premier 5000
Model/Catalog Number: 00024019255
Intelligent Quality Management 2 (iQM2) is used as the quality control and assessment system for the GEM Premier 5000 system.
Reason: BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
Root cause: Software Design Change
Recalling firm: Instrumentation LaboratoryType: Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood PhDistribution: US Nationwide distribution.Recall #Z-2821-2026