Baxter International · Model Travenol Flo-Gard 6100
Simple to operate. Simple to operate, inexpensive, with limited features, these heavy, unwieldy pumps (known as "boat anchors" by technicians) are popular with veterinary clinics because they are difficult for animals to overturn. They can also be used for human infusion; use standard Baxter tubing. Heavy internal transformer sometimes bends out of line, causing "Motor Defect" alarms. [b][i]Baxter has ended service on this product. Please see one of the available 3rd party companies for service and/or sales.[/i][/b]
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Occlusion alarm, no flow, upstream pressure high.
Root cause: Kinked set, clamped line, patient vein issue.
Symptoms: Air-in-line alarm, infusion paused.
Root cause: Air in set, improperly primed tubing.
Symptoms: Uncontrolled flow when door open.
Root cause: Worn roller or failed free-flow valve.
Symptoms: Stalled infusion, inaccurate rate.
Root cause: Peristaltic motor or encoder wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air detector / ultrasonic sensor | Periodic replacement | $$ | Moderate |
| Door microswitch | Periodic replacement | $ | Fairly easy |
| Occlusion sensor assembly | Periodic replacement | $$ | Moderate |
| IV administration set Consumable; per patient. | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
IV sets may leak.
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
Specific lots of Solution Sets with Duo-Vent Spike have the potential for inverted side clamps causing the tubing to be unusable.
The Surgical Tools and Cardiovascular Specialty marketing brochures distributed for Flo-Rester Disposable Internal Vessel Occluder, Flo-Thru Intraluminal Shunt, and Vascular Probe Intravascular Probe contain inaccurate content that does not align with the products Instructions for Use (IFU).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
There is a potential for foreign matter.
There is a discrepancy in the expiry date encoded into the 2D barcode. The correct expiration date is the human-readable date
Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
During production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.
The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.
Pumps submitted for service may not have been properly serviced before being returned to the customer.
Falsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
There is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
The mounting screws of the pole clamp may come loose, allowing the pole clamp assembly to detach from the rear pump housing, if they are not maintained on at least an annual basis.
There is an unsecure connection between the male Luer and the stopcock on the I.V. extension set.
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.