Abiomed, Inc. · Model Impella LD AIC Pump Set
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Intended use: A sterile tubular device with a built-in electrically-powered axial flow pump (e.g., a micro motor with an impeller) at the distal end, designed to provide circulatory assistance to the heart by pumping blood during heart failure. It is inserted through the lumen of the aorta and into the left ventricle, surgically or percutaneously via an artery, and connected to an external control unit; it moves blood from the left ventricle to the aorta across the aortic valve. It is typically used emergently post acute myocardial infarction (AMI), electively for high-risk cardiac catheterization, for open chest surgery, and during transport between hospitals. This is a single-use device.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.
Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.
A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential
Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.
Device packaged in incorrect outer box carton.
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.
A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Nine (9) Impella CP pumps failed inspection and were inadvertently released.
IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.
New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.
The following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature; Sort case tiles by AIC alarm color feature; Pump metric display on case tile feature
Pump not detected as connected to controller due to software issue.
A higher than expected rate of thrombus formation or deposition has been observed. The risk is for thrombus formations or deposits on indwelling central venous lines or cannulas may break free and enter into the Impella RP Flex, resulting in reduced flow, loss of support, or hemolysis
There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.
The Impella 5.5 pump experienced a heightened complaint rate for purge leaks.
Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Pump may separate at the motor housing while device is explanted